Nano-Flow Cytometry Provides LNP Size Measurements Closer to Cryo-TEM


How accurately are we really measuring LNP size?

A recent study in the Journal of Biomolecular Techniques compared four commonly used approaches—nano-flow cytometry, NTA, DLS, and cryo-TEM—for LNP size characterization.

The key finding was straightforward: under the conditions tested, NanoFCM and CytoFLEX nano produced LNP size measurements closer to cryo-TEM than NTA or DLS.The study, “A Rapid Method for Accurately Determining Lipid Nanoparticle Size Using Nano-Flow Cytometry,” highlights nano-flow cytometry as a promising option for rapid and accurate LNP size characterization.

1. Cryo-TEM Establishes a Direct Imaging Reference

Different LNP sizing technologies are based on different measurement principles, which can lead to variation in reported particle sizes.

To establish a direct reference, the researchers first used cryo-TEM to visualize the morphology of the LNPs and measure particle diameter directly.

The cryo-TEM analysis showed an average LNP diameter of 68.9 ± 16.3 nm, with most particles displaying a generally spherical morphology.

Because cryo-TEM directly visualizes individual particles, it provides a useful structural reference for comparing other sizing platforms.

This result also highlights an important point in LNP characterization: particle evaluation should not rely on a single mean size value. Particle size distribution and morphology should also be considered to obtain a more complete understanding of the formulation.

2. Four Platforms Compared in Parallel

After establishing the cryo-TEM reference, the same LNP batch was analyzed using four additional platforms:

  • NanoFCM
  • CytoFLEX nano
  • NanoSight NS300 (NTA)
  • Zetasizer Nano ZS (DLS)

The results showed that the particle sizes measured by NanoFCM and CytoFLEX nano were closer to the cryo-TEM result, while NTA and DLS reported larger particle sizes.

This difference may be related to the underlying measurement principles.

Both NTA and DLS typically report hydrodynamic diameter, which can be influenced by the particle’s surrounding solvent layer, sample conditions, data analysis methods, and particle distribution.

By contrast, nano-flow cytometry analyzes individual nanoparticles based on light scattering signals, enabling particle-by-particle measurement rather than relying on an ensemble average.

3. Statistical Analysis Further Supports the Nano-Flow Results

The researchers further compared the particle size results statistically.

The study found that particle sizes measured by NTA and DLS differed significantly from the cryo-TEM reference, while no significant difference was observed between cryo-TEM and either NanoFCM or CytoFLEX nano.

These results suggest that, for the LNP system evaluated in this study, nano-flow cytometry provided particle size measurements that more closely reflected the cryo-TEM reference.

 

4. Why Nano-Flow Cytometry Can Be Valuable for LNP Characterization

For LNP development, particle size is only one part of the characterization strategy.

From formulation screening to process optimization, researchers often need to evaluate multiple attributes, including:

  • Particle size and size distribution
  • Particle concentration
  • Surface protein modification
  • Protein-positive particle rate
  • Protein copy number
  • Nucleic acid loading
  • Free nucleic acid
  • Empty particle rate

One advantage of nano-flow cytometry is its ability to analyze individual nanoparticles, allowing both scattering and fluorescence signals to be measured at the single-particle level.

This means the same platform can potentially provide information not only on particle size, but also on surface modification and payload-related attributes, making it especially useful for targeted LNP and tLNP development.

 

Conclusion

This study provides a useful cross-platform comparison of LNP sizing methods.

Under the experimental conditions evaluated, nano-flow cytometry produced LNP size measurements closer to cryo-TEM than NTA or DLS, while also offering the advantage of rapid, single-particle analysis.

For LNP and tLNP development, this suggests that nano-flow cytometry can serve as a valuable complementary tool for particle characterization, particularly when researchers need to evaluate not only particle size, but also surface modification and nucleic acid loading at the individual-particle level.

As more LNP systems are evaluated and calibration and analysis workflows continue to improve, nano-flow cytometry may become an increasingly important platform for LNP formulation development and quality characterization.